Class B European Standard,
Weight: 64 (net:52)Kg,
Chamber size Ø: 240x360mm,
Capacity: 16 Liter,
Power: 1800W,
Voltage: 220/50V/Hz
16 Liter Autoclave, Class N
Adopting N-wclass (prED-13060), with thermal vacuumizing &
drying function, it can be used for A-type hollow instruments &
solid instraments.
Thermodynamic vacuum guarantees elimination of cold air
from the cavity and makes sure of perfect saturation of water
dream.
Three sterilizing water quantity selections are respectively for
different material, which embodies the individualized service
Automatic malfunction detecting system makes easy
maintaining.
* Liquid sterilizing option
16 Liter Autoclave, Class B European Standard
It adopts B-wclass(prED-13060), with 3-times pre-vacuum
preceding vacuum drying.
Residual humidity is <0.2: vacuum
measure-ment reaches-0.7bar.
It sterilizes and sanitizes miscellaneous packed or unpacked medical instruments, whether solid or A-type hollow, whether with multi- apertures or inset pipes.
It efficiently rid air from every hollow area of any
hollow instrument, such as hand-piece, making sure that water
steam reaches every corner.
It is installed with BOWIC & DICK which measures the
penetration of water steam.
Mini printer can be attached to record the process of sterilization
Distilled Water
Autoclave must adopt distilled water as operation medium.
Even Purified water has various minerals, and can produce furring scale after evaporation at high temperature, which, after a period of time, can lead to leaking of steam due to improper closure.
It can also lead to the clog in the narrow pipes and
pressure sensor, to the malfunction of temperature sensor and to other failures.
Once the furring scale enters handpiece and other instrument which are hollow inside or has apertures, it will clog the narrow pipes and axis, lowering the turning speed of handpiece, thus reducing its life span.
Therefore, using distilled water is necessary.